Icon NMR logo

IconNMR Automation Run Status:
Day Experiments: 00:00
Busy Until: Fri 11:51
Night Experiments: 00:00

Setup List
Holder Status Name ExpNo Experiment User Time Title
16 Completed 260123-iconnmr 1 N sst.1H.lsh 1H Lineshape: CHCl3 in d6-Acetone Specification result of humpcal: Linewidth at 0.55% of signal height ( < 7 Hz) Linewidth at 0.11% of signal height ( < 14 Hz) Halfwidth ( < 0.7 Hz) iconnmr 00:01:13 Supervisor Thomas
17 Completed 260123-iconnmr 2 N sst.1H.sen 1H Sensitivity 0.1% Ethyl Benzene in CDCl3 Specification result of sinocal: Best sino value found ( > 1050:1) iconnmr 00:01:02 Supervisor Thomas
18 Completed 260123-iconnmr 3 N sst.13C.sen 13C Sensitivity 10% Ethylbenzene in CDCl3 Specification result of sinocal: Best sino value found ( > 475:1) iconnmr 00:05:04 Supervisor Thomas
History List
Holder Time Name ExpNo Experiment User Load Atma Spin Lock Shim Acq Proc Title Remarks
18 09:45:33 260123-iconnmr 3 sst.13C.sen iconnmr + + - + + + + Supervisor Thomas Automation was stopped by user
17 09:33:27 260123-iconnmr 2 sst.1H.sen iconnmr + + - + + + + Supervisor Thomas
16 09:23:41 260123-iconnmr 1 sst.1H.lsh iconnmr + x - + + + + Supervisor Thomas ATM Failed, humpcal: Hump: at 0.11%: between 4822.8 and 4811.5 Hz = 11.3 Hz at 0.55%: between 4820.7 and 4814.8 Hz = 5.9 Hz at 50%: Half width at 4818.6 Hz = 0.56 Hz, humpcal: Hump: at 0.11%: between 4822.8 and 4811.5 Hz = 11.3 Hz at 0.55%: between 4820.7 and 4814.8 Hz = 5.9 Hz at 50%: Half width at 4818.6 Hz = 0.56 Hz
1 14:24:07 AutoCalibrate + - - - - + - AutoCalibrate: Success
1 14:22:31 AutoCalibrate x - - - - - - SampleJet: OP_CHANGER Command refused: System is in manual mode , AutoCalibrate - Sample could not be inserted
1 14:09:51 AutoCalibrate x - - - - - - SampleJet: OP_CHANGER Command refused: Error recovery required , AutoCalibrate - Sample could not be inserted